Section
5: SCPI command reference Series 2230G Triple-Channel Power Supplies
User's Manual
5-24 2230G-900-01 Rev. A / June 2018
MEASure[:SCALar]:CURRent[:DC]?
This command initiates and executes a new current measurement or queries the new measured current on a
specified channel or channels.
Group
Measurement
Syntax
MEASure[:SCALar]:CURRent[:DC]? {CH1|CH2|CH3|ALL}
Arguments
CH1|CH2|CH3|ALL The channel or channels on which to make a new current measurement.
Default value
Not applicable
Returns
<NR2> The newly measured current with the unit of measure A.
Example
MEAS:CURR? ALL
Makes a new current measurement on all channels.
See also
FETCh[:SCALar]:CURRent[:DC]? (on page 5-22)
MEASure[:SCALar]:POWer[:DC]?
This command initiates and executes a new power measurement or queries the new measured power.
Group
Measurement
Syntax
MEASure[:SCALar]:POWer[:DC]? {CH1|CH2|CH3|ALL}
Arguments
CH1|CH2|CH3|ALL The channel or channels on which to make a new power measurement.
Default value
Not applicable
Returns
<NR2> The newly measured power with the unit of measure W.

Table of Contents

Other manuals for Keithley 2230G-30-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2230G-30-3 and is the answer not in the manual?

Keithley 2230G-30-3 Specifications

General IconGeneral
BrandKeithley
Model2230G-30-3
CategoryPower Supply
LanguageEnglish

Related product manuals