Page #74 background imageLoading...
Page #74 background image
Section
5: SCPI command reference Series 2230G Triple-Channel Power Supplies
User's Manual
5-22 2230G-900-01 Rev. A / June 2018
MEASurement subsystem
The commands in the MEASurement subsystem are used to make measurements.
FETCh[:SCALar]:CURRent[:DC]?
This command queries the current reading on the specified channel.
Group
Measurement
Syntax
FETCh[:SCALar]:CURRent[:DC]? {CH1|CH2|CH3|ALL}
Arguments
CH1|CH2|CH3|ALL The selected channel or channels with the current readings to return.
Default value
Not applicable
Returns
<NR2> The newly current reading with the unit of measure A.
Example
FETC:CURR? ALL
Queries the current readings on all channels.
See also
MEASure[:SCALar]:CURRent[:DC]? (on page 5-24)
FETCh[:SCALar]:POWer[:DC]?
This command queries the present power measurement on a specified channel or channels.
Group
Measurement
Syntax
FETCh[:SCALar]:POWer[:DC]? {CH1|CH2|CH3|ALL}
Arguments
CH1|CH2|CH3|ALL The channel or channels with the measurement to return.
Default value
Not applicable

Table of Contents

Other manuals for Keithley 2230G-30-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2230G-30-3 and is the answer not in the manual?

Keithley 2230G-30-3 Specifications

General IconGeneral
BrandKeithley
Model2230G-30-3
CategoryPower Supply
LanguageEnglish

Related product manuals