Section
3: Function and features Series 2230G Triple-Channel Power Supplies
User's Manual
3-18 2230G-900-01 Rev. A / June 2018
Overtemperature protection
If the internal temperature of the power supply exceeds 85 ºC (185 ºF), the instrument beeps and
displays Over temperature, and then the instrument power turns off automatically.
Use the rear-panel terminals
When the input current of a DUT is high, there is a voltage drop on the wires between the DUT and
output terminal of the power supply. You can use remote sense wire connections to compensate for
the voltage drop and maximize measurement accuracy.
The rear-panel terminals are shown in the following figure.
Figure 11: Remote sense connections
The terminals labeled + and - are the positive and negative output terminals.
The terminals labeled S+ and S- are the positive and negative remote sense terminals.
There is a ground terminal for each channel on the rear panel. When a channel is enabled, its ground
terminal should be connected to a protective earth.

Table of Contents

Other manuals for Keithley 2230G-30-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2230G-30-3 and is the answer not in the manual?

Keithley 2230G-30-3 Specifications

General IconGeneral
BrandKeithley
Model2230G-30-3
CategoryPower Supply
LanguageEnglish

Related product manuals