Page #94 background image
Section
5: SCPI command reference Series 2230G Triple-Channel Power Supplies
User's Manual
5-42 2230G-900-01 Rev. A / June 2018
Example
STAT:OPER:INST?
Queries the Operation Instrument Event Register and then resets it.
See also
None
STATus:OPERation:INSTrument:ISUMmary<X>:CONDition?
This command queries the Operation Instrument Summary Condition Register of the status model for the
specified channel.
Group
Status
Syntax
STATus:OPERation:INSTrument:ISUMmary<X>:CONDition?
Arguments
<X> Channel number (1, 2, or 3)
Default value
0
Returns
<NR1> Range: 0 to 255
Example
STAT:OPER:INST:ISUM1:COND?
Queries the Operation Instrument Summary Condition Register of channel 1.
See also
None