EasyManuals Logo

Keithley 2230G-30-3 User Manual

Keithley 2230G-30-3
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
Section
5: SCPI command reference Series 2230G Triple-Channel Power Supplies
User's Manual
5-16 2230G-900-01 Rev. A / June 2018
Example
CAL:VOLT 30.002V
Sets the voltage value of the calibration point to 30.002 V.
See also
None
CALibrate:VOLTage:LEVel
This command sets the voltage calibration points.
Group
Calibration
Syntax
CALibrate:VOLTage:LEVel <point>
Arguments
<point> = P1|P2|P3|P4 Calibration point 1, 2, 3, or 4.
Points P1, P2, P3, and P4 must be calibrated in numeric order.
Default value
Not applicable
Returns
Not applicable
Example
CAL:VOLT:LEV P1
Sets the voltage calibration point to point 1.
See also
None

Table of Contents

Other manuals for Keithley 2230G-30-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2230G-30-3 and is the answer not in the manual?

Keithley 2230G-30-3 Specifications

General IconGeneral
BrandKeithley
Model2230G-30-3
CategoryPower Supply
LanguageEnglish

Related product manuals