EasyManuals Logo

Keithley 2230G-30-3 User Manual

Keithley 2230G-30-3
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
Section
5: SCPI command reference Series 2230G Triple-Channel Power Supplies
User's Manual
5-44 2230G-900-01 Rev. A / June 2018
Default value
0
Returns
<NR1> Range: 0 to 255
Example
STAT:OPER:INST:ISUM2?
Queries the Operation Instrument Summary Event Register of channel 2.
See also
None
STATus:PRESet
This command resets all bits in the status model.
Group
Status
Syntax
STATus:PRESet
Arguments
None
Default value
Not applicable
Returns
Not applicable
Example
STAT:PRES
Resets all the bits in the status model.
See also
None

Table of Contents

Other manuals for Keithley 2230G-30-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2230G-30-3 and is the answer not in the manual?

Keithley 2230G-30-3 Specifications

General IconGeneral
BrandKeithley
Model2230G-30-3
CategoryPower Supply
LanguageEnglish

Related product manuals