3-96 Keysight B1505A User’s Guide, Edition 12
Accessories
• Curve Tracer Test Adapter Socket Module (N1273A-013)
This module provides a socket available for connecting a test adapter designed
for connecting to Tektronix 370B/371B curve tracers. The socket module
internal connection is shown in Figure 3-33.
Figure 3-33 Curve Tracer Test Adapter Socket Module
To use this module, see the following simple instruction.
1. Connect the socket module to the Base/Gate, Collector/Drain, and
Emitter/Source connectors on the test fixture directly.
2. Connect your test adapter to the socket.
3. Set the DUT on your test adapter.
4. Close the fixture cover and perform measurement.
CAUTION Do not apply voltage/current over the maximum limit of the socket module.
Terminal related voltage: 3000 Vdc
Terminal related current: 100 mA
1
2
3
4
5
6
Base / Gate
Collector / Drain
Emitter / Source
AC/DC Guard
Internal connections
Socket Terminal N1273A Output
1: Collector/Drain Force Collector/Drain
2: Collector/Drain Sense No Connection
3: Emitter/Source Force Emitter/Source
4: Emitter/Source Sense No Connection
5: Base/Gate Force Base/Gate
6: Base/Gate Sense No Connection