Keysight M8070A Programming Guide 159
SCPI Command Reference 5
[:SOURce]:JITTer:HFRequency:BUNCorrelate[:STATe][?]
Syntax [:SOURce]:JITTer:HFRequency:BUNCorrelate[:STATe] ‘identifier’
<ON|OFF|1|0>
[:SOURce]:JITTer:HFRequency:BUNCorrelate[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', ‘M1.ClkOut’, or 'M*.DataOut'
<ON|OFF|1|0>: Enable/disable bounded uncorrelated jitter.
Return Range 1|0
Description Enables/disables the generation of bounded uncorrelated jitter at the
specified output.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :JITT:HFR:BUNC ‘M2.DataOut2’, off
[:SOURce]:JITTer:HFRequency:BUNCorrelate:AMPLitude[?]
Syntax [:SOURce]:JITTer:HFRequency:BUNCorrelate:AMPLitude ‘identifier’,
<NRf>
[:SOURce]:JITTer:HFRequency:BUNCorrelate:AMPLitude? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1' or 'M*.DataOut2', or 'M*.DataOut'
<NRf>: Set bounded uncorrelated jitter amplitude.
Return Range NRf
Description This command sets the bounded uncorrelated jitter amplitude in seconds
(TIME) or unit intervals (UINTerval). The units are set using the
[:SOURce]:JITTer:HFRequency:UNIT command.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :JITT:HFR:BUNC:AMPL ‘M2.DataOut2’, 400e-12