Page #165 background imageLoading...
Page #165 background image
Keysight M8070A Programming Guide 165
SCPI Command Reference 5
Input
Parameters
‘identifier’: 'M*.DataOut1' or 'M*.DataOut2', or ‘M*.DataOut’
<OFF|HP10>: Enable/disable high-pass filter.
Return Range OFF|HP10
Description This command enables/disables the high-pass filter for random jitter.
HP10 enables a 10 MHz high-pass filter.
This query returns the current settings.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :JITT:HFR:RAND:FILT:HPAS ‘M1.DataOut2’, off
[:SOURce]:JITTer:HFRequency:RANDom:FILTer:HPASs:VALue[?]
Syntax [:SOURce]:JITTer:HFRequency:RANDom:FILTer:HPASs ‘identifier’ <NRf>
[:SOURce]:JITTer:HFRequency:RANDom:FILTer:HPASs:VALue? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1' or 'M*.DataOut2', or ‘M*.DataOut’
<NRf>
Return Range 1 kHz to 10 MHz
The limit is dynamic, it depends on the low-pass setting.
Description This command controls the high pass filter value of the random jitter
source.
This SCPI is applicable for M8195A and M8196A.
This query returns the current settings.
Example :JITT:HFR:RAND:FILT:HPAss:VALue ‘M1.DataOut2’, 1E6.

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals