EasyManua.ls Logo

Keysight Technologies J-BERT M8020A - Page 167

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight M8070A Programming Guide 167
SCPI Command Reference 5
[:SOURce]:JITTer:HFRequency:SPECtrally[:STATe][?]
Syntax [:SOURce]:JITTer:HFRequency:SPECtrally[:STATe] ‘identifier’
<ON|OFF|1|0>
[:SOURce]:JITTer:HFRequency:SPECtrally[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', ‘M1.ClkOut’, or ‘M*.DataOut’
<ON|OFF|1|0>: Enable/disable spectrally distributed jitter.
Return Range 0|1
Description This command enables/disables the generation of spectrally distributed
random jitter.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example :JITT:HFR:SPEC ‘M2.DataOut2’, off
[:SOURce]:JITTer:HFRequency:SPECtrally:AMPLitude(*)[?]
Syntax [:SOURce]:JITTer:HFRequency:SPECtrally:AMPLitude(*) ‘identifier’ <NRf>
[:SOURce]:JITTer:HFRequency:SPECtrally:AMPLitude(*)? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', or ‘M*.DataOut’
<NRf>: Set spectrally distributed jitter amplitude.
Return Range 0 mUI to 78.71 mUI
Description This command sets the spectrally distributed jitter amplitude in unit
intervals or time.
A suffix attached to the :AMPLitude(*) command is used to address two
jitter amplitudes. The :AMPL1 <NR3> command changes the level of the
internal RJLF resource and the :AMPL2 <NR3> command changes the
level of the internal RJHF resource. Both levels are specified in rms.
The units are set using the [:SOURce]:JITTer:HFRequency:UNIT command.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example :JITT:HFR:SPEC:AMPL2 ‘M2.DataOut2’, 0.013