Keysight M8070A Programming Guide 255
SCPI Command Reference 5
:INPut:ALIGnment:EYE:TCEN
Syntax :INPut:ALIGnment:EYE:TCEN ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
Description This command initiates a search for the value of data/clock delay that puts
the active clock edge in the center of the data eye, midway between the
two relative delay points with a measured BER just in excess of the BER
configured by the :EYE:THReshold command. If successful, the command
leaves the data/clock delay at this value.
This SCPI is applicable for M8041A, M8051A and M8046A.
Example :INP:ALIG:EYE:TCEN ‘M1.DataIn1’
:INPut:ALIGnment:EYE:ABORt
Syntax :INPut:ALIGnment:EYE:ABORt ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
Description This command interrupts and aborts an alignment.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example :INP:ALIG:EYE:ABOR ‘M1.DataIn1’
:INPut:ALIGnment:EYE:THReshold[?]
Syntax :INPut:ALIGnment:EYE:THReshold ‘identifier’, <NRf>
:INPut:ALIGnment:EYE:THReshold? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
<NRf>: Set BER threshold.
Return Range 1E-9 to 1E-1
Description The command sets the BER threshold to be used in the determination of
the edges of the eye. The query returns the current BER threshold value.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example :INP:ALIG:EYE:THR ‘M1.DataIn1’, 1E-3

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Summary

Keysight M8000 Series of BER Test Solutions

Safety Summary

Covers general safety precautions, environment conditions, and handling instructions for instrument operation.

1 Programming Basics

Modular Configuration

Discusses the modular configuration of instruments for remote programming.

Using Identifiers

Explains the use of Location, Group Name, and Measurement Name identifiers.

2 Recommended Programming Techniques

Output Protection

Covers pattern generator output termination and protection algorithm.

Controlling the Output Levels

Explains how to set output levels using parameters like Vhigh, Vlow, Vamptl, Voffs.

Allowing the M8020A/M8030A/M8040A to Settle

Explains how instruments react to pattern changes and the time needed to settle.

3 Programming Examples

Initializing the Connection

Shows how to initialize the connection to the instrument using the ResourceName.

4 SCPI Command Language

SCPI Command Language - Introduction

Explains SCPI as the universal programming language for instrument control.

Important Points about SCPI

Covers key areas for SCPI usage: Instrument Model, Command Syntax, and Sending Commands.

Sending Commands to the M8020A/M8030A/M8040A

Details how to send commands, including an overview and using VISA.

5 SCPI Command Reference

Subsystems

Divides SCPI commands into subsystems reflecting instrument functionality.

PLUGin Subsystem

Describes the plugin interface for integrating C# assemblies.

Related product manuals