256 Keysight M8070A Programming Guide
5 SCPI Command Reference
:INPut:ALIGnment:EYE:RESult[:HEIGht]?
Syntax :INPut:ALIGnment:EYE:RESult[:HEIGht]? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
Description This is a query command that searches for the value of data amplitude that
puts the 0/1 threshold level midway between the upper and lower bounds
at which the error ratio exceeds the threshold value set by the
:INPut:ALIGnment:EYE:THReshold command.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example :INP:ALIG:EYE:RES? ‘M1.DataIn1’
:INPut:ALIGnment:EYE:RESult:WIDTh?
Syntax :INPut:ALIGnment:EYE:RESult:WIDTh? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
Description This is a query command that interrogates the eye width found by the most
recent search for the value of data/clock delay that put the active edge in
the center of the data eye.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example :INP:ALIG:EYE:RES:WIDT? ‘M1.DataIn1’
:INPut:ALIGnment:EYE:RESult:THReshold?
Syntax :INPut:ALIGnment:EYE:RESult:THReshold? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
Description Returns the threshold detected at the last alignment run.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example :INP:ALIG:EYE:RES:THR? ‘M1.DataIn1’