254 Keysight M8070A Programming Guide
5 SCPI Command Reference
:INPut:ALIGnment:EYE[:AUTO]
Syntax :INPut:ALIGnment:EYE[:AUTO] ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
Description This command starts an auto alignment.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
Example :INP:ALIG:EYE ‘M1.DataIn1’
:INPut:ALIGnment:EYE:ACEN
Syntax :INPut:ALIGnment:EYE:ACEN ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or M*DataIn
Description This command initiates a search for the 0/1 threshold voltage midway
between the two 0/1 threshold voltages with a measured BER just in
excess of the BER configured by the :EYE:THReshold command. If
successful, the command leaves the 0/1 threshold at this value.
This SCPI is applicable for M8041A, M8051A and M8046A.
Example :INP:ALIG:EYE:ACEN ‘M1.DataIn1’
:EYE:RESult:THReshold? :INPut:ALIGnment:EYE:RESult:THReshold?
on page 256
:EYE:RESult:DELay? :INPut:ALIGnment:EYE:RESult:DELay? on
page 257
:EYE:RESult:POLarity? :INPut:ALIGnment:EYE:RESult:POLarity? on
page 257