EasyManua.ls Logo

Keysight Technologies J-BERT M8020A

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight M8070A Programming Guide 267
SCPI Command Reference 5
This query returns the present state.
This SCPI is applicable for M8046A.
Example :N1000:CREC:LBWM "DCA1.SLOT1",RDEP
For detailed documentation, see :CRECovery{n}:LBWMode in the FlexDCA
online help.
:N1000:CRECovery:LOCKed?
Syntax :N1000:CRECovery:LOCKed? <identifier>
Input
Parameters
‘identifier’: ‘DCA*.SLOT*'
Description This query displays the clock recovery’s lock status.
Response: 1 (for locked) or 0 (for unlocked)
This SCPI is applicable for M8046A.
Example :N1000:CREC:LOCK? "DCA1.SLOT1"
For detailed documentation, see :CRECovery{n}:LOCKed the FlexDCA
online help.
:N1000:CRECovery:LSELect[?]
Syntax :N1000:CRECovery:LSELect <identifier>,{LOOP1 | LOOP2 | LOOP3 |
LOOP4}
:N1000:CRECovery:LSELect? <identifier>,
Input
Parameters
‘identifier’: ‘DCA*.SLOT*'
Description This command selects the type 2 loop transition frequency.
This query returns the present state.
This SCPI is applicable for M8046A.
Example :N1000:CREC:LSEL "DCA1.SLOT1",LOOP1
For detailed documentation, see :CRECovery{n}:LSELect in the FlexDCA
online help.

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Related product manuals