Keysight M8070A Programming Guide 269
SCPI Command Reference 5
:N1000:CRECovery:ODRatio:AUTO[?]
Syntax :N1000:CRECovery:ODRatio:AUTO <identifier>,{ON | OFF | 1 | 0}
:N1000:CRECovery:ODRatio:AUTO? <identifier>
Input
Parameters
‘identifier’: ‘DCA*.SLOT*'
Description This command enables or disables automatic selection of the recovered
clock’s output divide ratio.
This query returns the present state.
This SCPI is applicable for M8046A.
Example :N1000:CREC:ODR:AUTO "DCA1.SLOT1 ",OFF
For detailed documentation, see :CRECovery{n}:ODR:AUTO in the FlexDCA
online help.
:N1000:CRECovery:RDIVider[?]
Syntax :N1000:CRECovery:RDIVider[?] <identifier>,<ratio>
:N1000:CRECovery:RDIVider? <identifier>,
Input
Parameters
‘identifier’: ‘DCA*.SLOT*'
Description This command sets loop bandwidth symbol rate divider.
This query returns the present state.
This SCPI is applicable for M8046A.
Example :N1000:CREC:RDIV "DCA1.SLOT1 ",1667
For detailed documentation, see :CRECovery{n}:RDIVider in the FlexDCA
online help.
:N1000:CRECovery:RELock
Syntax :N1000:CRECovery:RELock <identifier>
Input
Parameters
‘identifier’: ‘DCA*.SLOT*'
Description This command initiates a re-lock of the clock recovery.