404 Keysight M8070A Programming Guide
5 SCPI Command Reference
threshold is specified with the
:PLUGin:EDIagram:EVALuation:BERThresh:THReshold[?] on page 404
command.
The query returns either 0 errors (ZERR) or BER threshold (BTHR).
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:EVAL:BERT 'MyMeasurement', BTHR
:PLUGin:EDIagram:EVALuation:BERThresh:THReshold[?]
Syntax :PLUGin:EDIagram:EVALuation:BERThreshold 'Identifier', <NRf>
:PLUGin:EDIagram:EVALuation:BERThreshold? 'Identifier'
Input
Parameters
Identifier: 'MyMeasurement'
Range 1E-12 to 1E-1
Description This command allows the user to specify BER Threshold value.
The query returns the current BER Threshold value.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:EVAL:BERT 'MyMeasurement', 1E-10
:PLUGin:EDIagram:EVALuation:TTIMe[?]
Syntax :PLUGin:EDIagram:EVALuation:TTIMe 'Identifier', <TEIG | TNIN>
:PLUGin:EDIagram:EVALuation:TTIMe? 'Identifier'
Input
Parameters
Identifier: 'MyMeasurement'
Range TEIG | TNIN
Description This command allows the user to specify the transition time. It can be
either 10/90 or 20/80.
The query returns the current setting.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:EDI:EVAL:TTIM 'MyMeasurement', TNIN