Triggered streaming requires the triggered dynamic sequence mode. In this mode, a
trigger is required for each time the sequencer is passing the start of the sequence.
Additional interruptions can be achieved by using the SINGLE or REPEATED mode for
other segments of the sequence. The advancement mode of the sequence itself needs
to be set to AUTO. The advancement mode of the last segment needs to be AUTO. In
triggered streaming, the instrument must be setup correctly to generate events with
triggers (refer to the section 8.9.1).
The following diagram illustrates the triggered streaming mode.