General Programming 8
M8190A User’s Guide 197
Command
:ARM:DEL 1E-13
Query
:ARM:DEL?
Set or query the fine delay settings. The unit is in seconds.
Table 8-10: Fine Delay Settings
Sample Frequency
Fine Delay Range
f
Sa
≥ 6.25 GSa/s
0 .. 30e-12
2.5 GSa/s ≤ f
Sa
< 6.25 GSa/s
0 .. 60e-12
f
Sa
< 2.5 GSa/s
0 .. 150e-12