3 Sequencing
96 M8190A User’s Guide
3.6.1.2 DYNAMIC CONTROL
The DYNAMIC CONTROL IN allows selecting a new segment or sequence at run time
from an external source. The connector provides 13 data bits, a select bit and a load
signal. As the internal sequencer table index is a 19 bit wide address the complete 19
bit have to be provided in two steps. The select signal is used to indicate which part of
the index is present at the data pins. The data is latched with the rising edge of the
Load signal.
DYNAMIC CONTROL IN Connector
Description: Receptacle, Mini D
Number of Contacts: 20
Manufacturer: 3M
Part Number: 10220-0210EC
Last Position
Position 2
Position1
Figure 3-4: DYNAMIC CONTROL IN connector