Sinusoid
Back to Arb Commands
Arb Type Current Voltage
Sinusoid
[SOURce:]ARB:CURRent:SINusoid:AMPLitude<value>, (@<chanlist>)
[SOURce:]ARB:CURRent:SINusoid:AMPLitude? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:AMPLitude <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:AMPLitude? [MIN|MAX,] (@<chanlist>)
Specifies the amplitude or peak value of the current or voltage sine wave. Referenced to I0 and V0 in
the Sinusoid diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:SINusoid:AMPLitude<value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:SINusoid:AMPLitude? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:SINusoid:AMPLitude <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:SINusoid:AMPLitude? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the amplitude or peak value of the current or voltage sine wave for Arb sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
The maximum value varies, depending on the currentor voltage
rating of the power module.
<amplitude>
Step 0 - 99, *RST 0 (none)
Programs a current amplitude: ARB:CURR:SIN:AMPL 5,(@1)
Programs a voltageamplitude: ARB:VOLT:SIN:AMPL 10,(@1)
Programs a current amplitude for step 0: ARB:SEQ:STEP:CURR:SIN:AMPL 5, 0,(@1)
[SOURce:]ARB:CURRent:SINusoid:FREQuency<value>, (@<chanlist>)
[SOURce:]ARB:CURRent:SINusoid:FREQuency? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:FREQuency <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:SINusoid:FREQuency? [MIN|MAX,] (@<chanlist>)
Specifies the frequency of the current or voltage sine wave in Hertz.