Back to Arb Commands
Arb Type Current Voltage
Trapezoid
[SOURce:]ARB:CURRent:TRAPezoid:END:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:TRAPezoid:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:END:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time in seconds, after the trapezoid, that the starting current or voltage level persists.
Referenced to t4 in the Trapezoid diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time in seconds, after the trapezoid, that the starting current or voltage level persists for
Arb sequence steps.
Parameter Typical Return
0 262.144 | MIN | MAX, *RST 0 <end time>
Step 0 - 99, *RST 0 (none)
Programs an end time after the current trapezoid: ARB:CURR:TRAP:END:TIM 2,(@1)
Programs an end time after the voltagetrapezoid: ARB:VOLT:TRAP:END:TIM 2,(@1)
Programs an end time for step 0:ARB:SEQ:STEP:CURR:TRAP:END:TIM 2, 0,(@1)
[SOURce:]ARB:CURRent:TRAPezoid:FTIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:TRAPezoid:FTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:FTIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:TRAPezoid:FTIMe? [MIN|MAX,] (@<chanlist>)
Specifies the fall time of the trapezoid in seconds. Referenced to t3 in the Trapezoid diagrams.
Keysight N6705C Operating and Service Guide 255
7 SCPI Programming Reference