SENSe:DLOG:TINTerval <value>
SENSe:DLOG:TINTerval? [MIN|MAX,]
Sets the integration time of the datalog measurement. Values from 1 millisecond to 60 seconds can
be specified
This command has been superseded by SENSe:DLOG:PERiod, but is still available for
backward compatibility.
Parameter Typical Return
0.0002048 to 60|MIN|MAX, *RST 0.1 <datalog period>
Specifies a datalog timeinterval of 0.01 seconds: SENS:DLOG:TINT 0.01, (@1)
l This function cannot be set to values less than 75 milliseconds when interleaved data logging is in
effect.
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer] <value>, (@<chanlist>)
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe[:UPPer] <value>, (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects the external datalog (Elog) current or voltage measurement range.๎œœ The value that you enter
must be the highest value that you expect to measure. Units are in amperes or volts.๎œœ The instrument
selects the range with the best resolution for the value entered.
Parameter Typical Return
0 to maximum|MIN|MAX, *RST MAX <maximum value of range>
Selects the 1 A range: SENS:ELOG:CURR:RANG 1, (@1)
Selects the 2 V range: SENS:ELOG:VOLT:RANG 2, (@1)
l When queried, the returned value is the maximum current or voltage that can be measured on the
range that is presently set.
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO? (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe:AUTO 0|OFF|1|ON , (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe:AUTO? (@<chanlist>)
Enables/disables seamless Elog measurement autoranging.
7 SCPI Programming Reference
320 Keysight N6705C Operating and Service Guide