Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enable seamless Elog measurement autoranging:
SENS:ELOG:CURR:RANG:AUTO ON, (@1)
SENS:ELOG:VOLT:RANG:AUTO ON, (@1)
l Autoranging does not include the 10 μA range, which must be selected individually.
SENSe:ELOG:FUNCtion:CURRent 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:CURRent? (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage? (@<chanlist>)
Enables or disables the Elog current or voltage measurement function.
Parameter Typical Return
0|OFF|1|ON, *RST currentOFF; voltage ON 0or1
Enables Elog currentmeasurements: SENS:ELOG:FUNC:CURR ON, (@1)
Enables Elog voltage measurements:SENS:ELOG:FUNC:VOLT ON, (@1)
SENSe:ELOG:FUNCtion:CURRent:MINMax 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:CURRent:MINMax? (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage:MINMax 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage:MINMax? (@<chanlist>)
Enables or disables logging of the minimum and maximum current or voltage values.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enables MIN/MAX Elog currentvalues: SENS:ELOG:FUNC:CURR:MINM ON, (@1)
Enables MIN/MAX Elog voltage values:SENS:ELOG:FUNC:VOLT:MINM ON, (@1)
SENSe:ELOG:PERiod <value>, (@<chanlist>)
SENSe:ELOG:PERiod? [MIN|MAX,] (@<chanlist>)
Sets the integration time of an Elog measurement.
Although the absolute minimum logging period is 102.4 microseconds, the actual minimum varies as a
function of the number of readings that are being logged (see Integration Period).
Parameter Typical Return
0.0001024 to 60|MIN|MAX, *RST 0.1 <period>
Specifies an Elog period of 0.01 seconds: SENS:ELOG:PER 0.01, (@1)
Keysight N6705C Operating and Service Guide 321
7 SCPI Programming Reference