SENSe:FUNCtion "CURRent"|"VOLTage"|"NONe", (@<chanlist>)
SENSe:FUNCtion? (@<chanlist>)
Selects a measurement function on models that do not have simultaneous voltage and current
measurement capability. This command is required so that the acquisition system knows which
function to acquire when a measurement is triggered. Selecting NONE disables the measurement
function.
Parameter Typical Return
"CURRent"|"VOLTage"|"NONE", *RST VOLTage CURR, VOLT, or NONE
Selects the current measurement function: SENS:FUNC "CURR", (@1)
l This command has been superceded by the SENSe:FUNCtion:CURRent and SENSe:FUNC-
tion:VOLTage commands, which should be used in new applications. SENSe:FUNCtion is still avail-
able for backward compatibility.
SENSe:FUNCtion:CURRent 0|OFF|1|ON, (@<chanlist>)
SENSe:FUNCtion:CURRent? (@<chanlist>)
SENSe:FUNCtion:VOLTage 0|OFF|1|ON, (@<chanlist>)
SENSe:FUNCtion:VOLTage? (@<chanlist>)
Enables or disables the current or voltage measurement function.
Parameter Typical Return
0|OFF|1|ON,
*RST current OFF; voltage ON
0or1
Enables currentmeasurements: SENS:FUNC:CURR ON, (@1)
Enables voltage measurements: SENS:FUNC:VOLT ON, (@1)
l These commands replace the previous SENSe:FUNCtion command and should be used in new
applications. SENSe:FUNCtion is still available for backward compatibility.
SENSe:FUNCtion:VOLTage:INPut MAIN|AUXiliary, (@<chanlist>)
SENSe:FUNCtion:VOLTage:INPut? (@<chanlist>)
Selects the voltage measurement input.
MAIN - selects the front panel binding posts as the measurement inputs.
AUXiliary - selects the rear panel auxiliary voltage inputs as the measurement inputs.
Parameter Typical Return
MAIN|AUXiliary, *RST MAIN MAIN or AUX
Selects the auxiliary voltageinput: SENS:FUNC:VOLT:INP AUX, (@1)
7 SCPI Programming Reference
322 Keysight N6705C Operating and Service Guide