The following trigger sources can be selected for the specified trigger system:
Trigger Source Description Trigger System
BUS Selects a remote interface trigger command. All trigger systems
CURRent<1-4> Selects the outputcurrent level of a channel. Acquire and DLOG triggers only
EXTernal Selects the rear panel Trigger In BNC connector and any connector
pins configured as trigger sources.
All trigger systems
IMMediate Generates a trigger as soon as the trigger system is INITiated. All trigger systems
PIN<1-7> Selects a digital portpin configured as a trigger input. All trigger systems except ARB and
DLOG
TRANsient<1-4> Selects the transient system as the trigger source. Acquireand Transienttriggers only
VOLTage<1-4> Selects the output voltage level of a channel. Acquire and DLOG triggers only
ARSK Selects the Arb Run/Stop key. DLOG triggers only
OOOK Selects the Output On/Off key. DLOG triggers only
Parameter Typical Return
BUS|CURRent<1-4>|EXTernal|PIN<1-7>|
TRANsient<1-4>|VOLTage<1-4>, ARSK, OOOK, *RST BUS
BUS, CURR<n>, EXT, PIN<n>, TRAN<n>, VOLT<n>, ARSK,
OOOK
Selects a current level on channel 1 as the acquire trigger:TRIG:ACQ:SOUR CURR1, (@1)
Selects the Trigger In BNC connector as the Arb trigger source:TRIG:ARB:SOUR EXT
Selects the Arb Run/Stop key as the DLOG trigger source:TRIG:DLOG:SOUR ARSK)
Selects a digital portpin 1 as the Elog trigger source: TRIG:ELOG:SOUR PIN1, (@1)
Selects the transient system of channel 4 as the trigger source: TRIG:TRAN:SOUR TRAN4, (@1)
Selects the voltagelevel of channel 3 as the acquisition trigger: TRIG:ACQ:SOUR VOLT3, (@1)
TRIGger:DLOG:CURRent[:LEVel] <value>, (@<chanlist>)
TRIGger:DLOG:CURRent[:LEVel]? [MIN|MAX,] (@<chanlist>)
TRIGger:DLOG:VOLTage[:LEVel] <value>, (@<chanlist>)
TRIGger:DLOG:VOLTage[:LEVel]? [MIN|MAX,] (@<chanlist>)
Sets the current or voltage triggered level of the data logger.๎ Applies when the data logger trigger
source is set to a level.๎ Values are specified in either amperes or volts.๎ The minimum and maximum
values depend on the ratings of the unit.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST 0
Voltage: 0 - maximum|MIN|MAX, *RST 0
The maximum values are 102% of the rating of the power module.
<triggered level>
Set the triggered current level to 3 A: TRIG:DLOG:CURR 3, (@1)
Set the triggered voltage level to 50 V: TRIG:DLOG:VOLT 50, (@1)
Keysight N6705C Operating and Service Guide 341
7 SCPI Programming Reference