[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude] <value> , (@<chanlist>)
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
[SOURce:]VOLTage:LIMit:NEGative[:IMMediate][:AMPLitude] <value>, (@<chanlist>)
[SOURce:]VOLTage:LIMit:NEGative[:IMMediate][:AMPLitude]? [MIN|MAX,] (@<chanlist>)
Sets the voltage limit when in current priority mode.๎œœ Units are in volts. The maximum value depends
on the voltage rating of the unit. The minimum value is the most negative value.
Parameter Typical Return
Positive: 0 to 102% of rating|MIN|MAX, *RST 102%
Negative: -102% of rating to 0|MIN|MAX, *RST -102%
<+ voltage limit>
<- voltagelimit>
Sets the positive voltagelimit to 20.4 V: VOLT:LIM 20, (@1)
Sets the negative voltage limitto -20.4 V: VOLT:LIM:NEG -20, (@1)
l MIN refers to the most negative value of the range. MAX refers to the least negative value of the
range.
l The query returns the programmed voltage limit in the form +n.nnnnnnE+nn for each channel spe-
cified. Multiple responses are separated by commas.
[SOURce:]VOLTage:LIMit:COUPle 0|OFF|1|ON, (@<chanlist>)
[SOURce:]VOLTage:LIMit:COUPle? (@<chanlist>)
Sets the voltage limit tracking state. The negative voltage limit tracks the positive voltage limit
setting.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0๎œœor๎œœ1
Enables the voltage limittracking state: VOLT:LIM:COUPON, (@1)
[SOURce:]VOLTage:MODE FIXed|STEP|LIST|ARB, (@<chanlist>)
[SOURce:]VOLTage:MODE? (@<chanlist>)
Sets the transient mode. This determines what happens to the output voltage when the transient
system is initiated and triggered.
FIXed - keeps the output voltage at its immediate value.
STEP - steps the output to the triggered level when a trigger occurs.
LIST - causes the output to follow the list values when a trigger occurs.
ARB - causes the output to follow the arbitrary waveform values when a trigger occurs.
7 SCPI Programming Reference