Low Range Voltage Programming - Models N6761A through N6766A
Step 1. Select the low-voltage programming range. The value entered must be the maximum voltage
of the range to be calibrated. This example selects the 5 V range of output 1.
Front Panel Menu Reference SCPI Command
From the Calibration menu, select Voltage, Low RangeProgramming. Then select
Start.
CAL:VOLT 5, (@1)
Step 2. Select the first voltage calibration point and enter the data. This value may be positive or
negative.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured dataEnter the data from the external DMM.
Press Selectwhen done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 3. Select the second voltage calibration point and enter the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P2 measured dataEnter the data from the external
DMM.Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Low Range Voltage Measurement - Models N6761A through N6766A
Step 1. Select the low-voltage measurement range. The value entered must be the maximum voltage
of the range to be calibrated. This example selects the 5 V range of output 1.
Front Panel Menu Reference SCPI Command
From the Calibration menu, select Voltage, Low RangeMeasurement. Then select
Start.
CAL:VOLT:MEAS 5, (@1)
Step 2. Select the first voltage calibration point and enter the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured dataEnter the data from the external DMM.
Press Selectwhen done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 3. Select the second voltage calibration point and enter the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P2 measured dataEnter the data from the external DMM.
Press Selectwhen done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Keysight N6705C Operating and Service Guide 393