3 VMA Mode
3.8 ACP Measurement
Couplings Coupled to "Scale Range" on page 1612 as follows:
Scale/Div = Scale Range/10 (number of divisions)
When "Auto Scaling" on page 1174 is ON, this value is automatically determined by the measurement
result
When you change a value, Auto Scaling automatically changes to Off
Preset 10.00 dB / Div
State Saved Saved in instrument state
Min 0.10 dB
Max 20 dB
Annotation <value> dB/ left upper of graph
Backwards
Compatibility
SCPI
:DISPlay:ACPower:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:PDIVision
Scale Range
Sets the Y-Axis scale range.
Remote Command Replace <meas> with the identifier for the current measurement
:DISPlay:<meas>:WINDow[1]:TRACe:Y[:SCALe]:RANGe <rel_ampl>
:DISPlay:<meas>:WINDow[1]:TRACe:Y[:SCALe]:RANGe?
Example
:DISP:CHP:WIND:TRAC:Y:RANG 100
:DISP:CHP:WIND:TRAC:Y:RANG?
Couplings Coupled to Scale/Div as follows
Scale Range = Scale/Div * 10 (number of divisions)
When you change this value, Auto Scaling automatically changes to OFF
Preset 100 dB
State Saved Saved in instrument state
Min 1
Max 200
Ref Position
Positions the reference level at the top, center, or bottom of the Y Scale display.
Changing the reference position does not change the reference level value.
Remote
Command
:DISPlay:ACPower:WINDow[1]:TRACe:Y[:SCALe]:RPOSition TOP | CENTer | BOTTom
:DISPlay:ACPower:WINDow[1]:TRACe:Y[:SCALe]:RPOSition?
Example
:DISP:ACP:WIND:TRAC:Y:RPOS CENT
1173 Vector Modulation Analyzer Mode User's &Programmer's Reference