3 VMA Mode
3.1 Digital Demod Measurement
If the trace is outside the bounds of some applicable limits, the text “Trace Fail”
will be displayed in red
If the window has no enabled limits, or the trace itself is not displayed, no
message is displayed for that trace
:CALCulate:DDEMod:LLINe[1]|2|3|4|5|6:FAIL? returns the trace limit testing
results.
:READ|FETCh|MEASure:DDEMod2? returns the scalar limit testing results.
Remote Command
:CALCulate:DDEMod:LIMit:STATe OFF | ON | 0 | 1
:CALCulate:DDEMod:LIMit:STATe?
Example
:CALC:DDEM:LIM:STAT OFF
:CALC:DDEM:LIM:STAT?
Preset
OFF
State Saved Saved in instrument state
Range
On|Off
Edit Scalar Limits
Enables you to edit the limit value for the scalar results.
Tx Power
Sets the limit for the transmit power pass/fail test. It fails when the measured power
is less than this limit value.
Remote Command
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:POWer <amp>
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:POWer?
Example
:CALC:DDEM:SEGM1:LIM:POW -30.0 dBm
Set the transmit power limit of segment1 to -30 dBm
:CALC:DDEM:SEGM1:LIM:POW?
query the transmit power limit of segment1
Preset -50.0
State Saved Yes
Min -200.0
Max 200.0
Backwards Compatibility
SCPI
:CALCulate:DDEMod:LIMit:POWer
375 Vector Modulation Analyzer Mode User's &Programmer's Reference