3 VMA Mode
3.1 Digital Demod Measurement
available with signal input RF” is returned
Preset
NORMal
State Saved Yes
Range Normal | Alternate
Spectrum Stitching
When Spectrum Stitching is turned on, multiple acquisitions overlapped in
frequency domain will be performed and then combined to get wide band IQ data.
This enables the instrument to capture and analysis wider signal, it may slow down
the measurement speed and it requires the signal to occupy the major part of the
analysis bandwidth.
Note: accurate trigger is needed to align multiple acquisitions in time domain, and
input signal need to be repeated and aligned with trigger precisely.
Remote
Command
[:SENSe]:DDEMod:STITching:SPECtrum[:STATe] ON | OFF | 1 | 0
[:SENSe]:DDEMod:STITching:SPECtrum[:STATe]?
Example
:DDEM:STIT:SPEC 1
:DDEM:STIT:SPEC?
Preset
OFF
State Saved Saved in instrument state
Cross-Correlated EVM
Enables Cross-Corelated EVM.
Remote Command
[:SENSe]:DDEMod:CCEVm[:ENABle] ON | OFF | 1 | 0
[:SENSe]:DDEMod:CCEVm[:ENABle]?
Example
:DDEM:CCEV 1
:DDEM:CCEV?
Preset
OFF
State Saved Saved in instrument state
Multi Channel Config
Lets you perform a detailed configuration of your component carriers, including
number of carriers, bandwidth, offset, integration bandwidth, and so on.
371 Vector Modulation Analyzer Mode User's &Programmer's Reference