EasyManua.ls Logo

Keysight N9010B - Operation Instrument Negative Transition; Operation Instrument Positive Transition

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Operation Instrument Negative Transition
Determines which bits in the "Operation Condition Query" on page 2586 Register
will set the corresponding bit in the "Operation Event Query" on page 2587 register
when the condition register bit has a negative transition (1 to 0).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:OPERation:INSTrument:NTRansition <integer>
:STATus:OPERation:INSTrument:NTRansition?
Example
:STAT:OPER:INST:NTR 1
Instrument Locked being cleared will be reported to the Instrument Summary of the Status Operation
register
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Operation Instrument Positive Transition
Determines which bits in the "Operation Condition Query" on page 2586 Register
will set the corresponding bit in the "Operation Event Query" on page 2587 register
when the condition register bit has a positive transition (0 to 1).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:OPERation:INSTrument:PTRansition <integer>
:STATus:OPERation:INSTrument:PTRansition?
Example
:STAT:OPER:INST:PTR 1
Instrument Locked being set will be reported to the Instrument Summary of the Status Operation
register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
2591 Vector Modulation Analyzer Mode User's &Programmer's Reference