Keysight N9010B - Detector Select Auto;Man; 3.5.10.5 Trace Function

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.5 Channel Power Measurement
AVERage with AVERage
POSitive (peak) with MAXHold
NEGative (peak) with MINHold
Preset
AVERage
State Saved Saved in instrument state
Range
NORMal|AVERage|POSitive|SAMPle|NEGative|RMS
Detector Select Auto/Man
Sets the Detector mode to Auto or Manual. In Auto, the proper detector is chosen
based on rules that take into account the measurement settings and other
instrument settings.
When you select any detector explicitly, this setting switches automatically to Man
(manual).
Remote
Command
[:SENSe]:CHPower:DETector:AUTO ON | OFF | 1 | 0
[:SENSe]:CHPower:DETector:AUTO?
Example
:CHP:DET:AUTO ON
:CHP:DET:AUTO?
Couplings When the Detector setting is Auto, switches to align with "Trace Type" on page 1737:
NORMal with Clear Write
AVERage with AVERage
POSitive (peak) with MAXHold
NEGative (peak) with MINHold
Preset
ON
State Saved Yes
3.5.10.5 Trace Function
Contains controls to:
Copy and Exchange traces
Preset or Clear all traces
929 Vector Modulation Analyzer Mode User's &Programmer's Reference