3 VMA Mode
3.1 Digital Demod Measurement
RMS EVM
Sets the limit for the RMS EVM or FSK Error measurement pass/fail test.
Remote
Command
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:EVM:RMS <real>
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:EVM:RMS?
Example
:CALC:DDEM:SEGM1:LIM:EVM:RMS 10.0
Set the RMS EVM limit of segment1 to 10 %
:CALC:DDEM:SEGM1:LIM:EVM:RMS?
query the RMS EVM limit of segment1
Notes The control label is “RMS EVM” for all modulation format except FSK formats
When modulation format is FSK formats, it is “FSK Error”
Preset 35.0
State Saved Yes
Min 0.00
Max 100.0
Backwards
Compatibility
SCPI
:CALCulate:DDEMod:LIMit:EVM:RMS
Frequency Error
Sets the limit for the frequency error measurement pass/fail test.
Remote
Command
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:FERRor[:HERTz] <real>
:CALCulate:DDEMod:SEGMent1|…|6:LIMit:FERRor[:HERTz]?
Example
:CALC:DDEM:SEGM1:LIM:FERR 40.0
Set the frequency error limit of segment1 to 40.0 Hz
:CALC:DDEM:SEGM1:LIM:FERR?
Query the frequency error limit of segment1
Preset 100.0
State Saved Yes
Min 0 Hz
Max 500 kHz
Backwards
Compatibility
SCPI
:CALCulate:DDEMod:LIMit:FERRor[:HERTz]