Keysight N9010B - 3.9.20 Trace; 3.9.20.1 Select Trace; 3.9.20.2 Trace Control

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.9 SEM Measurement
3.9.20 Trace
Lets you control the acquisition, display, storage, detection and manipulation of
trace data for the available traces. The Trace Control tab of this menu contains
radio-button selections for the trace type (Clear/Write, Trace Average, Max Hold,
Min Hold) and View/Blank setting for the selected trace.
3.9.20.1 Select Trace
Specifies the selected trace, which is the trace that will be affected when you
change trace settings.
Select Trace appears above the menu panel, indicating that it applies to all controls
in the menu panel. Select Trace is blanked if you select a tab whose controls do not
depend on the selected trace (for example, Trace Function).
Notes The selected trace is remembered even when not in the Trace menu
Dependencies For the Swept SA measurement:
In Image Suppress mode, when you select a trace it becomes the active trace, and the formerly
active trace goes into View
When you turn on Image Suppress, Update turns off for all traces except the selected trace
For the ACP measurement, when Meas Method is RBW, FAST or FPOWer, Select Trace is disabled
Preset Trace 1
State Saved Yes
3.9.20.2 Trace Control
The controls on this tab allow you to set the "Trace Type" on page 1737 and its
update mode.
There are four Trace Types:
Clear/Write
Trace Average
Max Hold
Min Hold
Each type handles data in a different way.
Vector Modulation Analyzer Mode User's &Programmer's Reference 1584