EasyManuals Logo

Keysight N9010B Programmer's Reference

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #245 background imageLoading...
Page #245 background image
3 VMA Mode
3.1 Digital Demod Measurement
"Pre-Adjust for Min Clipping" on page 241 selection is Elec Only.
Note that the Mech Atten value is not adjusted, and the value previously set is used.
Therefore, there is a case that IF Overload is still observed depending on the input
signal level and the Mech Atten setting.
Mech Atten Step
Controls the step size used when making adjustments to the input attenuation.
Labeled Mech Atten Step in Dual-Attenuator models and Atten Step in Single-
Attenuator models. In the Dual-Attenuator configuration, only affects the step size
of the mechanical attenuator.
Remote Command
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] 10 dB | 2 dB
245 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight N9010B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9010B and is the answer not in the manual?

Keysight N9010B Specifications

General IconGeneral
BrandKeysight
ModelN9010B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals