EasyManua.ls Logo

Keysight N9010B - Adjust IF Gain For; Spur Avoidance

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.2 Custom OFDM Measurement
Remote Command
[:SENSe]:OFDM:IF:GAIN:LEVel <rel_ampl>
[:SENSe]:OFDM:IF:GAIN:LEVel?
Example
:OFDM:IF:GAIN:LEV -10
:OFDM:IF:GAIN:LEV?
Notes Not available for B25 IF bandwidth option
Preset 0
State Saved Saved in instrument state
Min/Max -12/12
Adjust IF Gain For
Specifies the reference component carrier for adjustment of "IF Gain" on page 535.
Remote
Command
[:SENSe]:OFDM:IF:GAIN:REFerence:CCARrier CC0 | | CC7
[:SENSe]:OFDM:IF:GAIN:REFerence:CCARrier?
Example
:OFDM:IF:GAIN:REF:CCAR CC0
Preset CC0
State Saved Yes
Range CC0|CC1|CC2|CC3|CC4|CC5|CC6|CC7
Spur Avoidance
Option H1G has a spur at the -50 MHz offset of the Center Frequency, so this Spur
Avoidance function is provided to eliminate this spur.
Auto: move the spur completely out of analysis bandwidth, the trade-off is that
the available analysis bandwidth will be reduced to 400 MHz. And EVM could be
worse as the edge of the passband is used for analysis
Manual: move the spur by a specified value or turn off this feature by setting the
value to 0
Remote
Command
[:SENSe]:OFDM:CCARrier0|…|7:SAVoid:FREQuency <freq>
[:SENSe]:OFDM:CCARrier0|…|7:SAVoid:FREQuency?
[:SENSe]:OFDM:CCARrier0|…|7:SAVoid:AUTO ON | OFF | 1 | 0
[:SENSe]:OFDM:CCARrier0|…|7:SAVoid:AUTO?
Example
:OFDM:CCAR0:SAV:FREQ 50MHz
:OFDM:CCAR0:SAV:FREQ?
:OFDM:CCAR0:SAV:AUTO ON
:OFDM:CCAR0:SAV:AUTO?
537 Vector Modulation Analyzer Mode User's &Programmer's Reference