3 VMA Mode
3.1 Digital Demod Measurement
:DISPlay:DDEMod:WINDow[1]|2|…|6:X[:SCALe]:RLEVel?
Example
:DISP:DDEM:WIND3:X:RLEV 10
Set the X ref value of the third window to 10
:DISP:DDEM:WIND3:X:RLEV?
Query the X ref value of the third window
Couplings If Auto Scaling is On, the X Reference Value is determined by the trace data
Preset Depends on trace data
State Saved Yes
Min/Max -9.9E+37 / 9.9E+37
Ref Position
Determines the position from which the X scaling is calculated for the selected
trace. It can be set to the left side, center, or right side of the grid.
Remote
Command
:DISPlay:DDEMod:WINDow[1]|2|…|6:X[:SCALe]:RPOSition LEFT | CENTer | RIGHt
:DISPlay:DDEMod:WINDow[1]|2|…|6:X[:SCALe]:RPOSition?
Example
:DISP:DDEM:WIND3:X:RPOS CENT
Set the X ref position of the third window to CENT
:DISP:DDEM:WIND3:X:RPOS?
Query the X ref position of the third window
Preset Depends on trace data
State Saved Yes
Range Left | Ctr | Right
Auto Scaling
Toggles the Auto Scaling function On or Off.
Remote
Command
:DISPlay:DDEMod:WINDow[1]|2|…|6:X[:SCALe]:COUPle 0 | 1 | OFF | ON
:DISPlay:DDEMod:WINDow[1]|2|…|6:X[:SCALe]:COUPle?
Example
:DISP:DDEM:WIND3:X:COUP ON
Do the X auto scale for the third window
:DISP:DDEM:WIND3:X:COUP?
Query the X auto scale state of the third window
Couplings When this parameter is On, pressing the front-panel Restart key activates the scale coupling function,
that automatically determines scale per division and reference values based on the measurement
results. When you set a value to either X Width or Ref Value manually, Auto Scaling is automatically set
to Off
397 Vector Modulation Analyzer Mode User's &Programmer's Reference