3 VMA Mode
3.9 SEM Measurement
kHz, 100.0 kHz
:SEM:OFFS2:INN:LIST:BAND:VID?
:SEM:OFFS2:INN:LIST:BAND:VID:AUTO ON, ON, ON, ON, ON, ON
:SEM:OFFS2:INN:LIST:BAND:VID:AUTO?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS
Couplings When the Auto state is ON, Video BW is basically coupled with other parameters
Preset Automatically Calculated
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON | ON, ON, ON, ON, ON, ON, ON,
ON, ON, ON, ON, ON
State Saved Saved in instrument state
Saved in instrument state
Range Auto|Man
Min/Max 1 Hz/50 MHz
VBW/RBW
Selects the ratio between the video and resolution bandwidths.
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query for this parameter always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:BANDwidth:VIDeo:RATio <real>,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:BANDwidth:VIDeo:RATio?
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:BANDwidth:VIDeo:RATio:AUTO OFF | ON |
0 | 1,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:BANDwidth:VIDeo:RATio:AUTO?
Example
:SEM:OFFS2:INN:LIST:BAND:VID:RAT 0.1, 0.1, 0.1, 0.1, 0.1, 0.1
:SEM:OFFS2:INN:LIST:BAND:VID:RAT?
:SEM:OFFS:INN:LIST:BAND:VID:RAT:AUTO ON, ON, ON, ON, ON, ON
:SEM:OFFS:INN:LIST:BAND:VID:RAT:AUTO?
Notes Comma-separated list of values
OFFSet1 is for BTS, 2 for MS. Default is BTS
Couplings When Auto state is ON, the VBW/RBW is basically coupled with other parameters
Preset 0.01, 0.01, 0.01, 0.01, 0.01, 0.01, 0.01, 0.01, 0.01, 0.01, 0.01, 0.01 | 0.01, 0.01, 0.01, 0.01, 0.01, 0.01,
0.01, 0.01, 0.01, 0.01, 0.01, 0.01
OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF | OFF, OFF, OFF,
OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF
State Saved Saved in instrument state
1531 Vector Modulation Analyzer Mode User's &Programmer's Reference