EasyManua.ls Logo

Keysight N9010B - 3.2.5 Display; 3.2.5.1 Meas Display; Display Null Subcarriers; Display Data Subcarriers

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.2 Custom OFDM Measurement
3.2.5 Display
Opens the Display menu, which enables you to configure display items for the
current Mode, Measurement View or Window.
3.2.5.1 Meas Display
Contains controls for setting up the display for the current Measurement, View or
Window.
Display Null Subcarriers
Enables/disables display of null subcarriers in IQ traces. Null subcarriers have zero
amplitude; often this is just the DC subcarrier.
Remote Command
:DISPlay:OFDM:NULL:SUBCarrier OFF | ON | 0 | 1
:DISPlay:OFDM:NULL:SUBCarrier?
Example
:DISP:OFDM:NULL:SUBC 1
:DISP:OFDM:NULL:SUBC?
Preset OFF
State Saved Yes
Range OFF|ON
Display Data Subcarriers
Enables/disables display of data subcarriers in IQ traces.
Remote Command
:DISPlay:OFDM:DATA:SUBCarrier OFF | ON | 0 | 1
:DISPlay:OFDM:DATA:SUBCarrier?
Example
:DISP:OFDM:DATA:SUBC 1
:DISP:OFDM:DATA:SUBC?
Preset ON
State Saved Yes
Range OFF|ON
Display Pilot Subcarriers
Enables/disables display of pilot subcarriers in IQ traces.
465 Vector Modulation Analyzer Mode User's &Programmer's Reference