3 VMA Mode
3.9 SEM Measurement
When sending the command, missing values are not permitted; that is, if you want
to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STARt:SABSolute <real>,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STARt:SABSolute?
Example
:SEM:OFFS:INN:LIST:STAR:SABS -12.50 dBm, -12.50 dBm, -24.50 dBm, -11.50 dBm,
-11.50 dBm, -11.50 dBm
:SEM:OFFS:INN:LIST:STAR:SABS?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS
Preset 0dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0dBm, 0 dBm, 0 dBm|0 dBm, 0
dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm
State Saved Saved in instrument state
Min -200 dBm
Max 50 dBm
Abs2 Stop
Sets the 2nd absolute power level limit at the stop frequency for the selected inner
offset, ranging from –200 to +50 dBm. You can also toggle this function between
Couple and Manual. If set to Couple = ON, the Abs2 Stop power level limit is coupled
to "Abs2 Start" on page 1542, resulting in a flat limit line. If set to Man (Couple =
OFF), Abs2 Start and Abs2 Stop take different values, resulting in a sloped limit line.
The query returns values currently set to the offset stop 2nd absolute power limits.
When sending the command, missing values are not permitted; that is, if you want
to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:SABSolute <real>,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:SABSolute?
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:SABSolute:COUPle ON | OFF | 1 |
0,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:SABSolute:COUPle?
Example
:SEM:OFFS:INN:LIST:STOP:SABS -12.50 dBm, -24.50 dBm, -24.50 dBm, -11.50 dBm,
-11.50 dBm, -11.50 dBm
:SEM:OFFS:INN:LIST:STOP:SABS?
:SEM:OFFS:INN:LIST:STOP:SABS:COUP ON, ON, ON, ON, ON, ON
:SEM:OFFS:INN:LIST:STOP:SABS:COUP?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS