3 VMA Mode
3.9 SEM Measurement
State Saved Saved in instrument state
Range Integration BW|RRC Weighted
RRC Filter Alpha
Sets the alpha value for the RRC Filter.
Remote Command
[:SENSe]:SEMask:FILTer[:RRC]:ALPHa <real>
[:SENSe]:SEMask:FILTer[:RRC]:ALPHa?
Example
:SEM:FILT:ALPH 0.3
:SEM:FILT:ALPH?
Preset 0.22
State Saved Saved in instrument state
Min/Max 0.01/1.0
Non-Contiguous Meas Region
Selects the region to measure for the non-contiguous frequency allocation.
Option SCPI Comments
Outer
OUTer
Inner
INNer
Outer & Inner
OINNer
Available only in 5G NR and LTE-Advanced FDD/TDD Modes
Remote Command
[:SENSe]:SEMask:NCONtiguous:REGion INNer | OUTer | OINNer
[:SENSe]:SEMask:NCONtiguous:REGion?
Example
:SEM:NCON:REG INN
:SEM:NCON:REG?
Dependencies Available only in MSR, 5G NR and LTE-Advanced FDD/TDD Modes
OINNer is available only in 5G NR and LTE-Advanced FDD/TDD Modes
Preset
INNer
State Saved Yes
Range Inner | Outer | Outer & Inner
Sweep Type Rules
Selects which set of rules will be used for automatic selection of "Sweep Type" on
page 1504 when Sweep Type mode is Auto.
Vector Modulation Analyzer Mode User's &Programmer's Reference 1478