EasyManua.ls Logo

Keysight N9010B - Cumulate Mask Stop Frequency; Start Freq

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.9 SEM Measurement
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:CMASk[:STATe] ON | OFF | 0 | 1
[:SENSe]:SEMask:OFFSet[1]|2:INNer:CMASk[:STATe]?
Example
:SEM:OFFS:INN:CMAS 0
:SEM:OFFS:INN:CMAS?
Notes OFFSet 1 is for BTS, 2 for MS. Default is BTS
Preset 1|0
State Saved Yes
Range
ON|OFF
Cumulate Mask Stop Frequency
Specifies stop frequency of summing limit masks. For outside of the stop frequency,
the limit masks are not cumulated.
Remote Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:CMASk:FREQuency:STOP <freq>
[:SENSe]:SEMask:OFFSet[1]|2:INNer:CMASk:FREQuency:STOP?
Example
:SEM:OFFS:INN:CMAS:FREQ:STOP 500E6
:SEM:OFFS:INN:CMAS:FREQ:STOP?
Notes OFFSet 1 is for BTS, 2 for MS. Default is BTS
Dependencies Valid only when "Cumulate Mask " on page 1525 is ON
Preset 10.5 MHz
State Saved Yes
Min/Max 0 Hz/10 GHz
Start Freq
Specifies the start frequency for the currently selected offset. Also, enables you to
toggle that offset between On and Off. When sending the remote command, missing
values are not permitted; that is, if you want to change values 2 and 6, you must
send all values up to 6. Subsequent values remain unchanged. The query for this
parameter always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:FREQuency:STARt <freq>,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:FREQuency:STARt?
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STATe ON | OFF | 1 | 0,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STATe?
Example
:SEM:OFFS2:INN:LIST:FREQ:STAR 2.515 MHz, 2.715 MHz, 3.515 MHz, 4.00 MHz, 8.00
MHz, 12.50 MHz
:SEM:OFFS2:INN:LIST:FREQ:STAR?
Vector Modulation Analyzer Mode User's &Programmer's Reference 1526

Table of Contents

Other manuals for Keysight N9010B

Related product manuals