EasyManua.ls Logo

Keysight N9010B - Time; Spectrum; Demod; IQ Meas Time

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.1 Digital Demod Measurement
Sets the second window data to Search Time
Time
Time is the envelope of the data which is used to demodulation. It’s 20 % longer
than the Meas Interval, 10 % of the Meas Interval before the demodulated segment
and 10 % after.
Example
:DISP:DDEM:WIND2:DATA TIME
Sets the second window data to Time
Spectrum
Spectrum shows the FFT of the Raw Main Time waveform.
Example
:DISP:DDEM:WIND3:DATA SPEC
Sets the third window data to Spectrum
Demod
Displays the general demodulation results.
IQ Meas Time
Available for all modulation formats except 2-FSK.
IQ Meas Time is the measured time data results for the input signal. There are 6
available formats for this trace data: I-Q, Constellation, Real, Imaginary, I-Eye, Q-
Eye and Group Delay. Normally this trace data is displayed with I-Q format.
Format name Description
I-Q Real part of data is shown on horizontal axis, imaginary part is shown on vertical axis, Independent
variable (X axis) is normal to display
Constellation Same as I-Q, but for data with symbols defined, only the symbol points are shown as dots with no
connecting lines
Real (I) Real part of data is shown on a linear Y axis
Imaginary (Q) Imaginary part of data is shown on linear Y axis
I-Eye Real part of data is shown with X axis segmented (generally into 2 symbol segments) and each segment
is overlaid to show signal crossings at symbol boundaries
Q-Eye Same as I-eye but imaginary part of data is shown
Group Delay Shows the derivative of phase response with respect to time