3 VMA Mode
3.3 Monitor Spectrum Measurement
changes when these parameters are changed
When you manually set a value when in the Auto state, the state automatically changes to Man
Preset Automatically Calculated unless noted below
WCDMA Mode
Channel Power: 1.0 msOBW: 32.6 ms
ACP: 29 ms
State Saved Saved in instrument state
Min Other than non-sweeping hardware: Typically, 1 ms
Non-sweeping hardware: N/A
In the ACP measurement, when Meas Method is Fast Power, the minimum sweep time is span-
dependent and automatically calculated
Max Other than non-sweeping hardware: 4000 s
Non-sweeping hardware: N/A
Annotation The sweep time is displayed in the lower-right corner of the screen. The number of points is displayed
parenthetically, as:
Sweep 13.3 ms (1001 points)
A “#” mark appears before “Sweep” in the annotation when it is switched from Auto to Manual
coupling
Status Bits/OPC
dependencies
Meas Uncal is Bit 0 in the register:
STATus:QUEStionable:INTegrity:UNCalibrated
Auto Function
Remote
Command
[:SENSe]:<meas>:SWEep:TIME:AUTO OFF | ON | 0 | 1
[:SENSe]:<meas>:SWEep:TIME:AUTO?
Example Channel Power measurement:
:CHP:SWE:TIME:AUTO OFF
:CHP:SWE:TIME:AUTO?
Preset
WCDMA Mode
OFF
All others
ON
Minimum Acquisition Time
Available on non-sweeping hardware.
Specifies the minimum acquisition time for each “chunk” of the measurement result.
The instrument automatically divides Span into multiple chunks if needed.
Therefore, the total signal acquisition time for the entire Span is:
~(>~Minimum Acquisition Time) * (The number of chunks)
Vector Modulation Analyzer Mode User's &Programmer's Reference 650

Table of Contents

Other manuals for Keysight N9010B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9010B and is the answer not in the manual?

Keysight N9010B Specifications

General IconGeneral
BrandKeysight
ModelN9010B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals