EasyManua.ls Logo

Keysight N9010B - 7.3.19 CCDF Meas Results File Contents

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Save/Recall/Print
7.3 Save
Measurement Results consists of 20 comma-separated values, from MeasResult1
through MeasResult20.
MeasResult1 contains the same results as MEAS/READ/FETCh:SEMask1;
MeasResult2, MEAS/READ/FETCh:SEMask2; MeasResult3,
MEAS/READ/FETCh:SEMask3; and so on.
When imported into Microsoft Excel or a similar spreadsheet application, a typical
Meas Results file displays the header information above, followed by the data
section. A sample of the data rows appears below. Only the columns for
MeasResult1 through MeasResult6 are shown, due to lack of space:
MeasResul
t1
MeasResul
t2
MeasResul
t3
MeasResul
t4
MeasResul
t5
MeasResul
t6
…etc.
-999 0 -13 999 15.59025 -999
15.590253
59
0 -13 999 -999 -999
7.3.19 CCDF Meas Results File Contents
CCDF Meas Results Files are in CSV format, with a .csv extension. Each file
contains sets of measurement results, preceded by a header section. The header
section contains the following lines:
File ID string, which is MeasResult
Mode ID: Measurement ID, for example SA:PST
Firmware rev and model number
Option string
Automatic Trigger Time
Automatic Trigger Time State
CcdfCurrentCounts
Center Frequency
Center Frequency Step
Center Frequency Step State
Counts
Electrical Atten
2390 Vector Modulation Analyzer Mode User's &Programmer's Reference