3 VMA Mode
3.8 ACP Measurement
Mode: 5G NR
Remote
Command
[:SENSe]:ACPower:OFFSet[1]|2[:OUTer]:TYPE CTOCenter | CTOEdge | ETOCenter |
ETOEdge | RTOCenter | RTOEdge | RCTOCenter | SCTOCenter
[:SENSe]:ACPower:OFFSet[1]|2[:OUTer]:TYPE?
Subopcode: 1 = BTS/Downlink (Default), 2 = MS/Uplink
Example
:ACP:OFFS:TYPE ETOC
:ACP:OFFS:TYPE?
Preset
CTOCenter
State Saved Saved in instrument state
Range
CTOCenter | CTOEdge | ETOCenter | ETOEdge | RTOCenter | RTOEdge | RCTOCenter
| SCTOCenter
Diagrams for Offset Freq Define
Details depend on the selected mode.
Diagram for Modes other than MSR, LTEAFDD, LTEATDD, 5G NR