EasyManua.ls Logo

Keysight N9010B - Page 1297

Keysight N9010B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.8 ACP Measurement
:ACP:OFFS:TYPE?
Preset
CTOCenter
State Saved Saved in instrument state
Range
CTOCenter | CTOEdge | ETOCenter | ETOEdge | RTOCenter | RTOEdge | RCTOCenter
| SCTOCenter
Diagrams for Offset Freq Define
Details depend on the selected mode.
Diagram for Modes other than MSR, LTEAFDD, LTEATDD, 5G NR
1297 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight N9010B

Related product manuals