3 VMA Mode
3.8 ACP Measurement
retained and a new Trace Type command has been added. The :TRACe:MODE
command is retained for backwards compatibility, and the :TRACe:TYPE,
:TRACe:UPDate and :TRACe:DISPlay commands introduced for ongoing use. The
old Trace Modes are selected using :TRAC:MODE, whose parameters are mapped
into calls to :TRACe:TYPE, :TRACe:UPDate and :TRACe:DISPlay, and the old
global Averaging command [:SENSe]:AVERage[:STATe] is provided for
backwards compatibility. See the individual command descriptions for details.
When Average/Hold in the Meas Setup, Legacy Compatibility menu is ON, the
following is true for traces in Max Hold and Min Hold:
They ignore the Average/Hold number; Single for Max Hold causes one sweep
only, so switching to Single stops after the current sweep, and switching to Cont
starts again without clearing the accumulated result
Max Hold is not cleared on a Restart, Single or :INIT:IMM, but changing a
measurement parameter, like frequency or bandwidth etc., still restarts the Max
Hold
Preset
WRITe
State Saved The trace mode is an alias only
Backwards
Compatibility
SCPI
:TRACe[1]|2|…|6:MODE WRITe | MAXHold | MINHold | VIEW | BLANk
:TRACe[1]|2|…|6:MODE?
Backwards
Compatibility
Notes
The legacy :TRACe:MODE command is retained for backwards compatibility. In conjunction with the
legacy :AVErage command, it works as follows:
:AVErage ON|OFF sets/clears a variable that we will call average for the sake of this
discussion. This variable is maintained by the instrument solely for backwards compatibility. See
the [:SENSe]:AVERage[:STATe] command description below
:TRACe:MODE WRITe sets :TRACe:TYPE WRITe (Clear/Write) unless average is true, in
which case it sets it to :TRACe:TYPE AVErage. It also sets :TRACe:UPDate ON,
:TRACe:DISPlay ON, for the selected trace
:TRACe:MODE MAXHold sets :TRACe:TYPE MAXHold (Max Hold). It also sets
:TRACe:UPDate ON, :TRACe:DISPlay ON, for the selected trace
:TRACe:MODE MINHold sets :TRACe:TYPE MINHold (Min Hold). It also sets
:TRACe:UPDate ON, :TRACe:DISPlay ON, for the selected trace
:TRACe:MODE VIEW sets :TRACe:UPDate OFF, :TRACe:DISPlay ON, for the selected
trace
:TRACe:MODE BLANk sets :TRACe:UPDate OFF, :TRACe:DISPlay OFF, for the
selected trace
The query returns the same value as :TRACe:TYPE?, meaning that if you set :TRACe:MODE:VIEW
or :TRACe:MODE:BLANk, the query response will not be what you sent
Vector Modulation Analyzer Mode User's &Programmer's Reference 1370