3 VMA Mode
3.1 Digital Demod Measurement
Amplitude Ratio is the free decision distance in height of a digital signal and is
calculated as 2B/(A+B). A and B are the maximum and minimum amplitude
difference at the decision point. An ideal signal has a decision height of 100%.
Time Ratio is the free decision distance in width of a digital signal and is
calculated as 2B’/(A’+B’). A’ and B’ are the maximum and minimum zero-cross
time width. An ideal signal has a decision width of 100%.
Modulation Index Outer (SCPI only) is defined as the measure of extent of outer
amplitude variation about an un-modulated carrier. It is calculated as (VAmax
VAmin)/(VAmax + VAmin). VAmax is the largest amplitude at crest. VAmin is the
smallest amplitude at bottom. Maximum amplitude difference A is equal to
VAmax VAmin.
Modulation Index Inner (SCPI only) is defined as the measure of extent of inner
amplitude variation about an un-modulated carrier. It is calculated as (VBmax
VBmin)/(VBmax + VBmin). VBmax is the smallest amplitude at crest. VBmin is
the largest amplitude at bottom. Minimum amplitude difference B is equal to
VBmax VBmin.
Modulation On/Off Ratio (SCPI only) is defined as the ratio of 20log10((VAmax +
VBmax)/(VAmin +VBmin)).