The following table lists the bit definitions for the Standard Operation Register:
Bit Num-
ber
Bit Name
Decimal
Value
Definition
0 Calibrating 1 Instrument is performing a calibration.
1
Not Used
2 (Reserved for future use)
2
Not Used
4 (Reserved for future use)
3
Not Used
8 (Reserved for future use)
4 Measuring 16 Instrument is initiated, and is making or about to make a measurement.
5 Waiting for
Trigger
32 Instrument is waiting for a trigger.
6
Not Used
64 (Reserved for future use)
7
Not Used
128 (Reserved for future use)
8 Configuration
Change
256 Instrument configuration has been changed since the last INIT, READ? or MEASure?, either from
the front panel or from SCPI.
9 Memory
Threshold
512 Programmed number of measurements (DATA:POINts:EVENt:THReshold) have been stored in
measurement memory.
10 Instrument
Locked
1024 Set if a remote interface (GPIB, USB or LAN) has a lock (SYSTem:LOCK:REQuest?). Cleared when
the remote interface releases the lock (SYSTem:LOCK:RELease).
11
Not Used
2048 (Reserved for future use)
12
Not Used
4096 (Reserved for future use)
13 Global Error 8192 Set if any remote interface has an error in its error queue; cleared otherwise.
14
Not Used
16384 (Reserved for future use)
15
Not Used
32768 "0" is returned.
Command Summary
STATus:OPERation:CONDition?
STATus:OPERation:ENABle
STATus:OPERation[:EVENt]?
STATus:PRESet
STATus:QUEStionable:CONDition?
STATus:QUEStionable:ENABle
STATus:QUEStionable[:EVENt]?
SCPI Programming Reference
388
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals